The following information was provided by the initial reporter:
it was reported that anticancer drugs were administered via a c35-0 attached to a safe t cannula, which was selected for compatibility with the safe t port.Approximately 30 minutes after the start of administration, the patient exhibited abnormal symptoms.Upon investigation, it was determined that air was entering through the connection point and had entered the patient¿s bloodstream.Although appropriate measures prevented a serious outcome, this was an event that could have led to a life-threatening risk.The c35-0 attached to the safe-t cannula was connected at an angle.While an installation error is highly likely, the possibility of a defect in the c35-0 itself cannot be ruled out, so we would like this to be verified.Additional information received on 25 sep 2026
sales rep mentioned that:
it is unclear whether a standalone safe t-port is being connected to another manufacturer¿s iv set or whether an integrated iv set is being used.It¿s possible that the connection was made at an angle.This appears to be primarily a manual error, such as a misalignment during assembly.I didn't check myself, but it seems a nurse who noticed something was wrong checked it.It seems there was no leak, but apparently some air got in.The issue has not yet been resolved.Since there is a possibility of a product defect on the part of bd or cardinal, both companies have been asked to conduct a pir.Meanwhile, the hospital plans to provide training on how to use the device and implement measures to prevent future occurrences.
|