• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: B. Braun Medical Inc. Infusomat®; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B. Braun Medical Inc. Infusomat®; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Catalog Number 363419
Medical Device Problem Code Free or Unrestricted Flow (2945)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
As reported by the user facility: details of complaint (reported issue): "although the wheel of the 0.9% nacl tubing was completely closed, the blood flowed back up to the 0.9% nacl sac.The patient received approximately 50 ml of 0.9% nacl.It appears that the wheel does not allow complete occlusion of the tubing." no injury reported.
 
Additional Manufacturer Narrative
This report has been identified as b.Braun medical internal number (b)(4).The investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Infusomat®
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
B. Braun Medical Inc.
824 12th ave
bethlehem PA 18018
Manufacturer (Section G)
B. Braun Medical Inc.
824 12th ave
bethlehem PA 18018
Manufacturer Contact
elena reifenstein
824 12th ave
bethlehem, PA 18018
+1(833)425-1464
MDR Report Key50896572
Report Number2521402-2026-00893
Device Sequence Number10684683
Product Code FPA
UDI-Device Identifier04046964294570
UDI-Public(01)04046964294570(11)260312(17)290228(10)0062064624
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K142036
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/25/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number363419
Device Lot Number0062064624
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/15/2026
Initial Report FDA Received Date09/25/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/12/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1