| Catalog Number |
363419 |
| Medical Device Problem Code |
Free or Unrestricted Flow (2945)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
As reported by the user facility:
details of complaint (reported issue): "although the wheel of the 0.9% nacl tubing was completely closed, the blood flowed back up to the 0.9% nacl sac.The patient received approximately 50 ml of 0.9% nacl.It appears that the wheel does not allow complete occlusion of the tubing."
no injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This report has been identified as b.Braun medical internal number (b)(4).The investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.
|
| |
|