ABBOTT VASCULAR INC. MINI TREK¿ Coronary Dilatation Catheter; Catheters, Transluminal Coronary Angioplasty, Percutaneous Back to Search Results
Model Number 1012270-20
Medical Device Problem Codes Material Rupture (1546); Improper or Incorrect Procedure or Method (2017)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/01/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported the procedure was to treat a mildly tortuous and severely calcified lesion in the mid left circumflex artery.The 2.0x20mm rx mini trek balloon dilatation catheter was not soaked during the preparation for use.The device was advanced to the target lesion however during inflation, the balloon ruptured due to severe calcification of the lesion.Th procedure was completed using another device.There were no adverse patient effects.No additional information was provided.
 
Additional Manufacturer Narrative
H6 medical device problem code: 2017 - incorrect prep manufacturer's investigation is still pending at this time.Results and conclusions will be provided in the final report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MINI TREK¿ Coronary Dilatation Catheter
Common Device Name
Catheters, Transluminal Coronary Angioplasty, Percutaneous
Manufacturer (Section D)
ABBOTT VASCULAR INC.
3200 lakeside drive
santa clara CA 95054 -280
Manufacturer (Section G)
Abbott Vascular Costa Rica, Reg # 3009564766
52 calle 3 b31 coyol free zone
el coyol alajuela UNK
CS   UNK
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591--462
+1(951)914-5259
MDR Report Key50896585
Report Number2024168-2026-03713
Device Sequence Number21522857
Product Code LOX
UDI-Device Identifier08717648138164
UDI-Public(01)08717648138164(17)290131(10)60216G1
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K180040
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/25/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1012270-20
Device Catalogue Number1012270-20
Device Lot Number60216G1
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/09/2026
Initial Report FDA Received Date09/25/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/16/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-