| Catalog Number |
04745914001 |
| Medical Device Problem Code |
Non Reproducible Results (4029)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/09/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Roche diagnostics received a customer complaint regarding an allegedly discrepant sodium electrode result for a patient sample tested on a cobas c 501 module.The initial result was 180 mmol/l.The first repeat result was 140 mmol/l.The second repeat result, tested on another device, was 135 mmol/l.
|
| |
|
Additional Manufacturer Narrative
|
|
The sodium electrode lot number and expiration date were requested, but not provided.The field service engineer replaced the sample probe, calibrated the gear pump, and replaced rinse station tubing and aspiration tips.Controls and routine patient testing were performed; the outcome was positive.The instrument was fully functional.The investigation determined the service actions resolved the issue.
|
| |
|