• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Abbott Medical Regent Heart Valve with Flex Cuff; Heart-Valve, Mechanical

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Abbott Medical Regent Heart Valve with Flex Cuff; Heart-Valve, Mechanical Back to Search Results
Model Number 23AGFN-756
Medical Device Problem Code Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/31/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that on (b)(6)2026, a 23mm sjm regent heart valve w/flex cuff was chosen for a mechanical aortic valve replacement (mvr) procedure.After cardiac recurrence, it was found during esophageal ultrasound that the valve could not open and close normally, and the mechanical valve was removed.Therefore, a new a new 21mm sjm regent heart valve w/flex cuff was re-implanted while patient on bypass and esophageal ultrasound was performed, and the implanted valve opened and closed normally.There was no delay in the procedure and the patient was reported discharged.
 
Additional Manufacturer Narrative
The device is not returning for evaluation.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Regent Heart Valve with Flex Cuff
Common Device Name
Heart-Valve, Mechanical
Manufacturer (Section D)
Abbott Medical
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
Abbott Medical Reg# 2648612
lot 20 b st caguas
caguas 00725
PR   00725
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
+1(651)756-5400
MDR Report Key50896681
Report Number2135147-2026-06486
Device Sequence Number12219603
Product Code LWQ
UDI-Device Identifier05415067044174
UDI-Public(01)05415067044174(17)300521
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P810002
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/25/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number23AGFN-756
Device Catalogue Number23AGFN-756
Device Lot NumberC00030437
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/31/2026
Initial Report FDA Received Date09/25/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/22/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-