Customer reported that it was misreading.Medline received 1ea handheld aneroid in like-new condition, performed a visual inspection, and tested the sample by inflating it up to 300 mmhg; the test was repeated five times.During testing the cuff was able to inflate and deflate properly and the dial measured properly, and no signs of leaks or any other defects preventing the cuff and gauge to work as intended were observed.The complaint was not confirmed; the cause code description was not confirmed with received sample, and division has been informed and will continue to monitor trends.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|