| Model Number |
106531US |
| Medical Device Problem Code |
Electrical /Electronic Property Problem (1198)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/03/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the patient had driveline power fault.The log files were submitted for review and the modular cable was exchanged.The alarm resolved after system controller was exchanged.There was no evident erosion at the modular cable site.The event log file review showed dates from
(b)(6) 2026.The event log file captured driveline power faults (power b) starting from the beginning of the data and continued for the remainder of the log file.The picture provided showed the inline connector to be clean with no notable debris or corrosion.
|
| |
|
Additional Manufacturer Narrative
|
|
No further information was provided.A supplemental report will be reported once the manufacturer¿s investigation is completed.
|
| |
|
Search Alerts/Recalls
|