The account reported finding adhesive issue dressing, and the complaint stated the suresite dressing was failing in almost every way.A sample was received; medline received 1 unopened case of kit,iv start,chloraprep applicator, evaluated 16 ea of the kits, and only one kit had a dressing where the paper frame separated from the dressing itself so it was no longer able to be used; 15 ea of the dressings tested peeled correctly.Finished good production records and finished good pack build were reviewed, with no run without approvals, deviations or non-conformance reports that might have had an impact on this issue and no issues with the build that likely contributed to the reported issue.Per the investigation and sample received, the reported issue was confirmed; the cause was identified as deriving from the raw materials process when the force between frame liner and the film is too weak, and medline is working with the component manufacturing facility for further evaluation and corrective actions.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|