Respironics, Inc. Respironics; Ventilator, Continuous, Minimal Ventilatory Support, Facility Use
|
Back to Search Results |
|
| Model Number |
V60 V60Plus Ventilator |
| Medical Device Problem Code |
Excessive Heating (4030)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/10/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Philips received a complaint by the customer on the v60 indicating a device malfunction as the device had a high temperature blower motor error.It was reported that there was no patient involvement at the time the issue was discovered.No patient or user harm was reported.Based on the case information provided, the device did not meet product specifications.The alleged malfunction impacted the user¿s ability to use the device properly.The customer called technical support to report a "blower temperature high" alarm error.No accessories, including third-party devices, were being used that may have contributed to the issue.The customer checked the event log but was unable to verify the error in the event log.After powering on the device in normal operational mode and letting it run for over 30 minutes, the customer did not discover any alarms.The remote service engineer (rse) informed the customer of the recommended repair per the v60 service manual if the high blower temperature error reoccurs.The customer noted that further testing will be performed.This investigation is ongoing.
|
| |
|