• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Abbott Medical Amplatzer Amulet; Cardiac Plug

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Abbott Medical Amplatzer Amulet; Cardiac Plug Back to Search Results
Model Number 9-ACP2-010-025
Medical Device Problem Code Patient-Device Incompatibility (2682)
Health Effect - Clinical Codes Perforation (2001); Pericardial Effusion (3271)
Date of Event 08/31/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that on (b)(6) 2026, a 25mm amplatzer amulet left atrial appendage occluder was chosen for a procedure.The patient lost pressure once device released.A pericardiocentesis was performed and drained 750cc of blood from the pericardium.After draining 750cc patient became stable and pressures normalized.Next day, the physician mentioned, the patient ended up being sent to surgery and the device was extracted and their appendage was over-sewn.The cardiothoracic (ct) surgeon mentioned to physician that there was some trauma to the distal end of the appendage, and one of the stabilizing wires perforated the appendage.The patient was in the intensive care unit (icu) and recovering.
 
Additional Manufacturer Narrative
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Amplatzer Amulet
Common Device Name
Cardiac Plug
Manufacturer (Section D)
Abbott Medical
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
St. Jude Medical #3014918977
edificio #44 calle 0, ave. 2
el coyol alajuela 1897- 4050
CS   1897-4050
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
+1(651)756-5400
MDR Report Key50896799
Report Number2135147-2026-06488
Device Sequence Number18812512
Product Code NGV
UDI-Device Identifier00811806013497
UDI-Public(01)00811806013497(17)310131(10)11255406
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P200049
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/25/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number9-ACP2-010-025
Device Catalogue Number9-ACP2-010-025
Device Lot Number11255406
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/31/2026
Initial Report FDA Received Date09/25/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/10/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-