The account reported finding holes n glove w burn marks.A photo and a sample were received for investigation; production records found no nonconformances noted by production or quality control personnel during first article, in-process, or case packaging inspections, and sample evaluation found a hole was visible in one of the fingers of the glove.The reported issue was confirmed.The vendor stated that the most probable cause might have been due to foreign particles on the surface of the former which was not removed in the washing tank, causing deposit in the glove; centurion's complaint history and production data over the past five years were reviewed with no other reports of a hole in the glove identified, and the issue appears to be isolated.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|