• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Medline Industries, LP MEDLINE; TRANSPORT CHAIR, KNOCKDOWN, STEEL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Medline Industries, LP MEDLINE; TRANSPORT CHAIR, KNOCKDOWN, STEEL Back to Search Results
Catalog Number MDS808200KD
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/29/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that "when customer was trying to remove the right legrest he noticed the welding on the top clip/hook to attach it to the unit is broken off.The welding on it snapped off.".
 
Additional Manufacturer Narrative
The customer reported a damaged footrest, and that when customer was trying to remove the right leg rest, he noticed the welding on the top clip/hook to attach it to the unit is broken off.A physical sample was not provided; a photo depicting the defect was provided and only a visual review of the defect could be conducted, and the product appears to be in used condition.The investigation results were confirmed, and after reviewing the photos it was discovered that the upper pin plate of the right-side leg rest has fractured off of the leg rest.Given the evidence provided in the complaint and photos, the root cause of this issue is unknown; the division will monitor through quarterly trending.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
TRANSPORT CHAIR, KNOCKDOWN, STEEL
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
madelyn ruppert
three lakes drive
northfield, IL 60093--275
MDR Report Key50896977
Report Number1417592-2026-50782
Device Sequence Number12262334
Product Code INM
UDI-Device Identifier00888277329983
UDI-Public00888277329983
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/24/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberMDS808200KD
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/29/2024
Initial Report FDA Received Date09/25/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-