• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Nihon Kohden Corporation Life Scope VS; Bedside Monitor

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Nihon Kohden Corporation Life Scope VS; Bedside Monitor Back to Search Results
Model Number BSM-3532
Medical Device Problem Codes Device Alarm System (1012); Protective Measures Problem (3015)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/28/2026
Type of Reportable Event Death
Event or Problem Description
The biomedical engineer (bme) reported a patient event involving a death while the patient was connected to the bedside monitor (bsm).The bme requested a review of the device logs to determine whether the monitor alarmed appropriately and whether a staff member silenced the alarm.
 
Additional Manufacturer Narrative
The biomedical engineer (bme) reported a patient event involving a death while the patient was connected to the bedside monitor (bsm).The bme requested a review of the device logs to determine whether the monitor alarmed appropriately and whether a staff member silenced the alarm.The event occurred on (b)(6) 2026 from 7:36 a.M.To 7:48 a.M.Logs were obtained from the bsm and the central nurse station (cns; pu-681ra, sn: (b)(6)).An nk clinical application specialist (cas) reviewed the logs and spoke with the bme and risk manager.The review confirmed that the monitor alarmed appropriately.The bme and risk manager discovered that a provider at the cns had silenced the crisis alarms; as a result, the nurses were unaware that the patient was coding.The customer acknowledged that the event involved user error.Nihon kohden continues to investigate the reported event.Nihon kohden will submit a supplemental report in accordance with 21 cfr section 803.56 when additional information becomes available.Additional device information: d10 concomitant medical device: the following device was used in conjunction with the bsm: cns: model #: pu-681ra serial #: (b)(6) device manufacturer data: 08/01/2019 (aug) unique identifier (udi) #: (b)(4) returned to nihon kohden: na.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Life Scope VS
Common Device Name
Bedside Monitor
Manufacturer (Section D)
Nihon Kohden Corporation
attn: shama mooman
1-31-4 nishiochia
shinjuku-ku, tokyo 161-8 560
JA  161-8560
Manufacturer (Section G)
Nihon Kohden Tomioka Corporation
attn: shama mooman
1-1 tajino
tomioka city, gunma 370-2 314
JA   370-2314
Manufacturer Contact
shama mooman
seibu bldg 2, 4th floor 1-11-2
kusunokidai tokorozawa, saitama 359-8-580
JA   359-8580
MDR Report Key50897036
Report Number8030229-2026-07179
Device Sequence Number6026507
Product Code MHX
UDI-Device Identifier04931921113660
UDI-Public(01)04931921113660
Combination Product (Y/N)N
Initial Reporter StateNE
Initial Reporter CountryUS
PMA/510(K) Number
K201949
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date (Section B) 09/25/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberBSM-3532
Device Catalogue NumberBSM-3532A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/28/2026
Initial Report FDA Received Date09/25/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/02/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CNS
Outcome Attributed to Adverse Event Death;
Patient Age22 YR
Patient SexFemale
Patient Weight169 LB
Patient RaceWhite
-
-