| Model Number |
2120 |
| Medical Device Problem Code |
Inaccurate Flow Rate (1249)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/28/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during testing, the failed the downstream occlusion (dso) pressure test, with a measured pressure value of 8.47 psi.There was no patient involvement associated with this event, and no patient harm was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The suspect pump was returned for evaluation.Review of the event history log (ehl) identified no events related to the reported issue.A review of the 12-month service history confirmed that the device had not undergone servicing during this period.Visual inspection revealed a torn downstream occlusion (dso) seal and a dented upstream occlusion (uso) seal.Functional testing was performed to evaluate the reported failure of the downstream occlusion (dso) pressure test; however, the issue could not be replicated.The dso sensor and uso seal were replaced as part of the repair.The probable cause of the reported issue could not be determined.Following repair, the device passed all functional testing.
|
| |
|