According to the customer, when prepping the manifold the syringe would "unscrew itself from the manifold" and the incident "can" introduce air into the system; the customer did not report patient involvement, and the issue was noted "before the case started." a review of the device history records was performed for the indicated packaging and component lots, and the returned samples were unopened; the suspect manifold and syringe were used in an attempt to replicate the same defect of the complaint.The connected samples were properly engaged and determined to pass, exhibiting no damage to either component, or the complaint is not confirmed.The root cause remains inconclusive at this time; no correction is needed at this time, no additional corrective actions to be taken by medline at this time, and medline will continue to monitor this anomaly for additional occurrences.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|