Consumer reported, the plunger rod was difficult to move on two of the syringes.Stated, he had to push really hard to get the insulin to come out and that caused him pain in the injection site.Stated, 2 more syringes leaked around the part where the needle sits because it was bent.Stated, the needle was not bent.The part that holds the needle was bent.So that's four syringes in total he had a problem with.Lot: 4065022,
catalog: 328468,
date of event: unknown,
samples: yes, cl.This intake record was reviewed as part of capas #pr-00078 and #pr-00080 and deficiencies were identified.The following actions were taken one additional complaint record was created for plunger difficult to move, one complaint record for needle pain, one complaint record for leakage, and one complaint record for hub damaged/bent.With these actions completed, this intake record is now consistent with applicable procedures (b)(4), and regulatory requirements, and separate complaints were opened for each separate event.There was appropriate follow-up (if this was required), and each event is addressed based on the most current information.No further action is required at this time unless new information becomes available.(b)(6), (b)(6) 2026.
|