According to the customer, the towels "leave lint behind when used to clean the surgeons¿ hands and the area surrounding the patient during surgery," and the customer reported that there was no patient involvement and that lint fibers did not enter the body cavity.The provided photo and videos were reviewed, the vendor was notified of the issue, and trend analysis was performed by reviewing the past two years of complaint data.Lint particulates could be seen shedding from the or towels and sticking to the customers hand, confirming the reported issue and that the product failed to meet specifications.The vendor noted that the root cause of the linting issue was attributed to residual short cotton fibers, particulate buildup on equipment surfaces, and insufficient visual inspection; no further investigation into the cause and complaint is warranted.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|