| Model Number |
4471 |
| Medical Device Problem Codes |
Failure to Capture (1081); High Capture Threshold (3266)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/10/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the pacemaker and this right ventricular (rv) lead right ventricular automatic threshold (rvat) test was suspended due to a loss of capture.Also, technical services (ts) suspected this pacemaker stored inappropriate ventricular episodes due to an atrial arrythmia with rapid ventricular response (rvr).Ts also noted the last rvat was above programmed outputs and was trending upwards.Ts discussed the loss of capture with the health care professional (hcp).The hcp planned to consult with cardiology.This rv lead remains in service.No adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This product investigation is still in progress.This report will be updated when product investigation is complete.
|
| |
|