It was reported that when using the bd pyxis¿ medbank tower, a user was unable to dispense oxycodone ir 5 mg tablet.Upon investigation, it was determined that the medication had been destocked by the system, making it unavailable for removal from the station.The issue was related to the item's inventory status.The customer stated that there was a delay in patient care.There were no adverse events or injuries reported based on this event.
|
|
A review of the complaint history for sn (b)(6)was performed in salesforce which did not locate similar complaint(s) with the same failure mode for this serial number.A review of the device history record for sn (b)(6)was performed from the date of manufacture, 30-apr-2024 and confirmed that this device was not previously returned for servicing and there were no production failures which correlates to the customer reported issue.Upon investigation of the actual device used in this incident, it was determined that the item was destocked by the system and all drawers had failed, leaving the station unresponsive.The field service engineer (fse) arrived on site and waited for access keys before performing further inspection.During troubleshooting, a loose universal serial bus port was identified on the main logic board, causing intermittent loss of communication with the drawers when the connection was disturbed.After obtaining access, the controller board was replaced, and medbank support was contacted to configure the new board.Following configuration, drawer functionality was verified using diagnostic software, and all drawers operated normally with testing completed successfully.The system functioned as intended after field service engineer repaired the device.
|