The customer reported that the technician cleaning scopes in the scope room complained of getting wet through the gown, specifically at the seam on the sleeves, and was not getting wet on the trunk.Medline received samples and reviewed them with the division; the samples were reviewed and the wrong material was used on the arms of the gowns.The reported problem was confirmed.This was identified as a manufacturing issue, with cause listed as mfg: assembly/human error, and a stock check was initiated, product was put on hold, a recall was issued, and a scar was issued to the vendor.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|