• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDLINE; ASPIRATOR,SUCTION,VAC-ASSIST

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDLINE; ASPIRATOR,SUCTION,VAC-ASSIST Back to Search Results
Catalog Number HCS7000
Medical Device Problem Codes Corroded (1131); Power Problem (3010); Suction Failure (4039)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/13/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that "will not turn on.".
 
Additional Manufacturer Narrative
The account reported finding the sample will not turn on.Medline quality received one unit, visually inspected the sample and observed the aspirator was used, attempted to turn the unit on for functional evaluation, and opened the device to check for internal damage and defects; the unit turn on with a low humming sound, no suction pressure was detected, and heavy rust and fluid damage inside the back compartment indicated fluid within the unit.The specific issue of unit not turn on was confirmed as the shaft, eccentric, and the bearing are unable to operate due to fluid damage with received sample.No trends were identified, and the complaint review process has been completed.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
ASPIRATOR,SUCTION,VAC-ASSIST
Manufacturer Contact
vamshi thupakula
three lakes drive
northfield, IL 60093--275
MDR Report Key50897515
Report Number1417592-2026-50861
Device Sequence Number14928550
Product Code JCX
UDI-Device Identifier40080196769293
UDI-Public40080196769293
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K042349
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/24/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberHCS7000
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer03/09/2026
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/13/2026
Initial Report FDA Received Date09/25/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-