The customer reported issue "not sticking" and "sticking too well" with optifoam gentle ex, sacrum, 9x9; per the facility, "patients are reporting pain with dressing removal and epidermis peeling off when peeling back the dressing," and the dressings were placed as a preventative measure and were being removed to check the skin underneath.Upon sample review, there were no abnormalities with the adhesion and silicone adhesive, and inspection records for this product were reviewed and the adhesion results met specifications.The customer reported issue could not be confirmed with the received sample.The cause was documented as not confirmed with received sample, no adverse trending identified, and this complaint is ok to close.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|