The customer reported a concern of melting therahoney hd sheets, with liquid pooling out on the sides of the therahoney sheet and honey melted off the sheet.A sample was provided for investigation, and the supplier was notified of the issue.The honey pooling out of the sides of the dressing was confirmed with the sample provided for investigation.It appears that the dressing was either overfilled or melted, or a combination of both; root cause may be attributed to production at the supplier as they may have overfilled this lot of dressings.No adverse trending was identified, and the complaint was ok to close.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|