| Model Number |
72404127 |
| Medical Device Problem Codes |
Difficult or Delayed Activation (2577); Material Deformation (2976); Migration (4003); Air/Gas in Device (4062)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/16/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that this artificial urinary sphincter (aus) was difficult to activate at the activation appointment.Per the patient they had a fall before the appointment.A surgical procedure was performed during which the entire device was explanted and replaced.Upon explant the physician discovered that the pump had migrated and tubing was potentially twisted.It was also identified there was air in the pump, but the pressure regulating balloon (prb) was still fully inflated with fluid.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Model number/catalog number: 72404132.Serial number: n/a.Batch/lot number: 1100929081.Model/catalog description: belt cuff fgs 5.0 cm iz.Unique identifier (udi) #: (b)(4).Model number/catalog number: 72400024.Serial number: n/a.Batch/lot number: 1100968790.Model/catalog description: balloon fgs 61-70 cm.Unique identifier (udi) #: (b)(4).
|
| |
|