Model Number 480422-03
Medical Device Problem Codes Failure to Deliver Energy (1211); Failure to Seal (4070)
Health Effect - Clinical Code Hemorrhage/Blood Loss/Bleeding (1888)
Date of Event 09/02/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during a da vinci-assisted total gastrectomy procedure, after sealing with the vessel sealer extend (vse) instrument, unexpected bleeding was observed from the omentum.The amount of blood loss was not provided.No blood transfusions were administered, and the bleeding was controlled using a fenestrated bipolar forceps (fbf) instrument.The procedure was completed.
 
Additional Manufacturer Narrative
Intuitive surgical, inc.(isi) did not receive a da vinci product to perform failure analysis.Vessel sealer logs showed this vessel sealer extend (vse) instrument was installed 6 times and passed homing 6 times.130 cut complete events were noted.The instrument was used with an e-200 generator and logs show one error: "e-200 energy activation halted by an instrument or instrument cable connected to the blue upper bipolar port on the e-200 generator while sealing with a vessel sealer instrument.¿ the instrument was used for approximately 74 minutes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Endowrist
Common Device Name
VESSEL SEALER EXTEND
Manufacturer (Section D)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA 95051
Manufacturer (Section G)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA 95051
Manufacturer Contact
izabel nielson
3410 central expressway
santa clara, CA 95051
+1(408)523-2100
MDR Report Key50897744
Report Number2955842-2026-39861
Device Sequence Number6001541
Product Code NAY
UDI-Device Identifier10886874115661
UDI-Public(01)10886874115661(10)K17260130
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K173337
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Other,Notification
Type of Report Initial
Report Date (Section B) 09/25/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Operator of Device Health Professional
Device Model Number480422-03
Device Catalogue Number480422
Device Lot NumberK17260130
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 09/02/2026
Initial Report FDA Received Date09/25/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/29/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
ISIFA2022-01-C
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
da Vinci Instruments and Accessories.
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-