| Model Number |
480422-03 |
| Medical Device Problem Codes |
Failure to Deliver Energy (1211); Failure to Seal (4070)
|
| Health Effect - Clinical Code |
Hemorrhage/Blood Loss/Bleeding (1888)
|
| Date of Event |
09/02/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that during a da vinci-assisted total gastrectomy procedure, after sealing with the vessel sealer extend (vse) instrument, unexpected bleeding was observed from the omentum.The amount of blood loss was not provided.No blood transfusions were administered, and the bleeding was controlled using a fenestrated bipolar forceps (fbf) instrument.The procedure was completed.
|
| |
|
Additional Manufacturer Narrative
|
|
Intuitive surgical, inc.(isi) did not receive a da vinci product to perform failure analysis.Vessel sealer logs showed this vessel sealer extend (vse) instrument was installed 6 times and passed homing 6 times.130 cut complete events were noted.The instrument was used with an e-200 generator and logs show one error: "e-200 energy activation halted by an instrument or instrument cable connected to the blue upper bipolar port on the e-200 generator while sealing with a vessel sealer instrument.¿ the instrument was used for approximately 74 minutes.
|
| |
|