| Model Number |
2110 |
| Medical Device Problem Code |
Device Sensing Problem (2917)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/13/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that, during testing, the device alarmed indicating that the cassette was partially unlatched after the cassette had been inserted, the door closed, and the cassette locked.During investigation testing, a defective latch lock sensor was confirmed.There was no patient involvement associated with this event, and no patient harm was reported.However, if this issue were to recur during infusion, the pump may not operate as intended and therapy could be interrupted.
|
| |
|
Additional Manufacturer Narrative
|
|
The suspect device was returned for evaluation.Review of the event history log (ehl) identified a "cassette partially unlatched" error in the event history.A review of the device records found no available 12-month service history.Visual inspection revealed no anomalies related to the reported complaint.Functional testing, including a latch lock test, confirmed a faulty latch sensor, thereby confirming the reported issue.The latch sensor was replaced as part of the repair.The probable cause of the reported issue was determined to be a faulty cassette latch sensor.Following repair, the device passed all functional testing.
|
| |
|