• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Abbott Medical Amplatzer Amulet; Cardiac Plug

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Abbott Medical Amplatzer Amulet; Cardiac Plug Back to Search Results
Model Number 9-ACP2-010-025
Medical Device Problem Code Patient Device Interaction Problem (4001)
Health Effect - Clinical Codes Stroke/CVA (1770); Thrombosis/Thrombus (4440)
Date of Event 09/03/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that on (b)(6) 2026, a 25mm amplatzer amulet left atrial appendage occluder was implanted in the patient.During procedure, the patient's activated clotting time (act) levels were 250-274s and 14,000 units of initial heparin dose, then additional 3000 heparin administered after the first act resulted as 250.Post procedure, 45 day follow up transesophageal echocardiogram (tee) showed a 1.5x1.0cm device related thrombus (drt).The patient was non-compliant with taking aspirin and plavix, and completely did not take any of her dual antiplatelet therapy (dapt) medication.The patient was placed on warfarin after the drt was found.Then, over (b)(6) the patient presented to another hospital with a stroke event.The drt was noted to be smaller but still present at that time.The patient remains on warfarin due to the ability to monitor international normalized ratio (inr) levels closely for medication non-compliance.The patient was reported stable.
 
Additional Manufacturer Narrative
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Amplatzer Amulet
Common Device Name
Cardiac Plug
Manufacturer (Section D)
Abbott Medical
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
St. Jude Medical #3014918977
edificio #44 calle 0, ave. 2
el coyol alajuela 1897- 4050
CS   1897-4050
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
+1(651)756-5400
MDR Report Key50897836
Report Number2135147-2026-06505
Device Sequence Number10758678
Product Code NGV
UDI-Device Identifier00811806013497
UDI-Public(01)00811806013497(17)290930(10)10555412
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
P200049
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Type of Report Initial
Report Date (Section B) 09/25/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number9-ACP2-010-025
Device Catalogue Number9-ACP2-010-025
Device Lot Number10555412
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/03/2026
Initial Report FDA Received Date09/25/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/09/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-