| Model Number |
35700BAX2 |
| Medical Device Problem Code |
False Alarm (1013)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/03/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that a spectrum pump had an air in line sensor failure during programming/setup.There was no patient involvement.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
E1 initial reporter phone no: (b)(6).The device was received for evaluation.During functional testing, the customer reported issue of ¿air in line sensor failure¿ was reproduced.A review of the event history log revealed "air in line'¿ messages.A service history review revealed no indication that the parts replaced during servicing caused or contributed to the reported event.The reported condition was verified.The cause of the condition was determined to be ultrasonic sensor out of specification.The ultrasonic sensor will be replaced to address this issue.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|