| Model Number |
LVP-0004 |
| Medical Device Problem Code |
No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/10/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The following has been reported: pump not connected to the network and password not working to login.A preliminary review of the device logs could not be performed.The device was returned for evaluation.The log files indicate a processor hardware failure (linux core)/som failure.The som module will be removed and replaced during the repair process.During evaluation, the (right/left/upper) loading pin assembly was found to have the spring positioned beyond the retaining c-clip.The affected loading pin(s) shall be replaced as part of the repair process.An active infusion was not stopped.Reporting as a conservative measure for the linux core issue identified during investigation.No adverse effects were reported.
|
| |
|