Manufacturer's investigation conclusion:
the reported event of fluid ingress on the backup battery (serial number: (b)(6)) was confirmed via customer submitted photo.The customer submitted photo showed evidence of fluid ingress in the backup battery compartment and on the backup battery.The submitted log files showed no indication that the fluid ingress prevented the system controller from supporting the system in the data reviewed.The root cause of the fluid ingress was not able to be conclusively determined via this investigation.Battery inspection data was reviewed for the 11v backup battery, serial number (b)(6), revealing that the battery passed all inspection testing.The heartmate 3 left ventricular assist device (lvas) instructions for use (ifu), and the heartmate 3 patient handbook, are currently available.The current revision of the instructions for use (ifu) can be found on the eifu page of the abbott website.Section 10 of the patient handbook, ¿safety checklists¿, instructs users to regularly inspect their equipment, including their system controllers, and to avoid using equipment that appears damaged.Users are encouraged to replace any equipment that appears damaged.The heartmate 3 patient handbook, section titled "emergency contact list" cautions users to call their hospital contacts if they think, for any reason, any portion of their equipment is not functioning as usual, is broken, or they are uncomfortable with the operation of the equipment.No further information was provided.The manufacturer is closing the file on this event.
|