| Catalog Number |
08837058190 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/27/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
We received an allegation about discrepant results for 2 patients' samples tested with elecsys anti-hcv ii assay on a cobas e 801 analytical unit.Patient 1:
initial result: 1.62 coi (reactive).1st repeat result: 1.65 coi (reactive).2nd repeat result: 1.65 coi (reactive).On (b)(6)2026, the sample was repeated at a different customer's branch:
3rd repeat result: non-reactive (tested using the electrochemiluminescence immunoassay (eclia) method).The result was confirmed by an enzyme-linked fluorescent assay (elfa) method.Patient 2 was tested on (b)(6)2026:
initial result: 2.75 coi (reactive).1st repeat result: 2.80 coi (reactive).2nd repeat result: 2.78 coi (reactive).On (b)(6)2026:
the result from another laboratory, which uses the chemiluminescence immunoassay (clia) method, was 0.11 s/co (negative).
|
| |
|
Additional Manufacturer Narrative
|
|
The cobas e 801 analytical unit serial number was (b)(6).The investigation is ongoing.
|
| |
|