According to the customer, the iv set "failed" due to "valves leaking" requiring the set to be replaced; the customer did not report any occurrence of adverse event related to the administration of medications, serious injury, medical intervention, or follow up care related to the reported incident.Functional inspection was conducted with the provided sample; medline tested the sample and hooked the tubing up to an iv bag, and the tubing was not leaking and functioned normally.The reported concern could not be confirmed, as no issues have been found, and the cause code description was not confirmed with received sample.The division will continue to monitor this issue for trending, and there has been no change to the reportability decision associated with this complaint.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|