| Model Number |
PED2-300-14 |
| Medical Device Problem Codes |
Material Deformation (2976); Activation Failure (3270)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/23/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Medtronic received information regarding a pipeline that failed to open at the distal end and was damaged.The patient was undergoing a procedure for flow diversion treatment of a left middle cerebral artery (mca) unruptured saccular aneurysm.The aneurysm max diameter was 5.4mm and the neck diameter was 4mm.The distal landing zone was 2mm; the proximal landing zone was 2.6mm.Patient's vessel tortuosity was normal.It was reported that the pipeline and all accessory devices were prepared as indicated in the instructions for use (ifu).During dep loyment, the pipeline stent tip failed to open.The pipeline was resheathed, and deployment was reattempted.When the catheter was retracted to the middle of the pipeline for deployment but the tip still failed to open.The stent was resheathed and removed.When removed, damage to the tip of the pipeline stent was observed.The pipeline was replaced with another of the same model which was delivery and deployed to successfully complete the procedure.The stent showed good wall apposition with stasis in the aneurysm.There was no harm or injury to the patient.Ancillary devices: 6f long sheath, navien 5f 125cm intermediate catheter, phenom 27 microcatheter.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|