Information was received from a patient who was receiving morphine via an implantable pump for non-malignant pain.It was reported they were finally able to get the communicator on after probably 100 attempts to get it to turn on even though it was fully charged.Then the communicator shut off.Since then, they had only been able to connect one time, so twice total.The patient charged the communicator all the way and would get the green light turned on when plugged into the wall, but when they pressed the power button, it did not turn on.They held the button, but that did not work.They could not give a bolus and had severe problems with everything.The issue was not resolved.A request was sent to the repair department to replace the communicator.
|
|
Continuation of d10: product id: tm90t0, lot# serial#: (b)(6), product type: accessory, product id: a820, lot# serial#: unknown, product type: software.Section d information references the main component of the system.Other relevant device(s) are: product id: tm90t0, serial/lot#: (b)(6), ubd: 27-feb-2027, udi#: (b)(4), h3: analysis of the communicator "[s/n: (b)(6) ]" found that the micro usb charge port was loose, the device passed the bench test, and the device failed the final functional jbal test.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|