| Catalog Number |
51-109060 |
| Medical Device Problem Code |
Tear, Rip or Hole in Device Packaging (2385)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/22/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during inspection of circulated stock inventory, a sterile packaging pouch was found damaged.There was no patient involvement.Attempts have been made and no further information has been provided.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4)
g2: foreign - event occurred in japan.Visual examination of the returned cartons confirmed item and lot label identification.Blisters were opened to evaluate the pouch and pre-tear.All devices have creasing and tearing on the pouches at the pre-tear.No other damage was noted during visual evaluation.Pouch seal widths were measured and were less than the acceptable 6mm.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.Reported event did not occur in an operating room or as part of a medical procedure.The root cause of the reported event can be attributed to transit damage and a packaging design issue.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Search Alerts/Recalls
|