| Model Number |
G22692 |
| Medical Device Problem Code |
Premature Activation (1484)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/12/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
During preparation for an endoscopic variceal ligation in the esophagus, the physician selected a cook 6 shooter saeed multi-band ligator.It was reported that the user opened the package and discovered that the bands had premature released.The user changed to another device to complete the procedure.This occurred prior to patient contact; there was no impact to the patient.In addition, the patient sustained no clinical consequence and there were no adverse effects to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Continued:
section e: phone: (b)(6).Investigation evaluation:
the product said to be involved was returned in a biohazard bag with an open box and tray from the lot number provided in the report.The label matches the product returned.One band was not included in the return.Our laboratory evaluation of the product said to be involved confirmed the report.Only 3 bands remained on the barrel as received and the remaining bands had moved slightly along the barrel as in premature deployment.Two bands returned loose in the tray.A visual examination of the device was performed.The trigger cord was examined and all twelve deployment beads were present.The beads could not be verified for correct location due to the trigger cord still being attached to the barrel.The beads were examined using magnification and found to be correctly filled and had no evidence of excess flash.The trigger cord was intact and not broken.The length of the trigger cord was measured between the knots which is within the established tolerance.An evaluation of the handle wheel movement was performed, and the wheel functioned as intended.A product-specific discrepancy that could have caused or contributed to this observation was not observed during our laboratory analysis.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion:
our evaluation of the returned device confirmed the report.A definitive cause for this observation could not be determined because the actual prior to use product preparation and handling conditions could not be duplicated in the laboratory setting.This limits our ability to conclusively determine a cause.Premature band deployment can occur if the device experiences excessive pressure during general handling or shipment.Prior to distribution, all 6 shooter saeed multi-band ligators are subjected to a visual inspection to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action:
based on the quality engineering risk assessment no corrective action is warranted at this time.A review of the complaint history was conducted.Based on this review, the likelihood of this type of report is rare.Quality assurance will continue to monitor for complaint trends.
|
| |
|