| Model Number |
2110 |
| Medical Device Problem Code |
Failure to Infuse (2340)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/01/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the pump displayed a "cassette was not attached properly" error.Patient involvement associated with this event is unknown.However, if the issue were to recur during an infusion, the pump may not function as intended and could result in an interruption of therapy.
|
| |
|
Additional Manufacturer Narrative
|
|
B3 ¿ date of event: the exact date of the event is unknown; therefore, the first day based on the icu medical awareness date was entered as the event date.The suspected device was returned for evaluation.There was no 12-month service history during the review.Visual inspection had been carried out, and no anomalies were observed related to the complaint.Functional tests were performed, and complainant allegation could be confirmed and replicated.Replaced downstream occlusion (dso) sensor.The probable cause of the reported issue is defective dso sensor.The device passed all functional testing after repair.
|
| |
|