| Model Number |
URF-P7 |
| Medical Device Problem Code |
Environmental Particulates (2930)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/25/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported that the uretero-reno videoscope had a separate issue.During the initial evaluation of the device, it was observed that dirt was on the objective lens.No patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
The final investigation has been completed.The failure identified during the investigation was confirmed.The root cause could not be identified.However, the most probable cause of this complaint is expected or random component failure without any design or manufacturing issue.Olympus will continue to monitor the field performance of this device.Should any additional relevant information become available, a supplemental report will be submitted.
|
| |
|