The customer reported that when they have to push saline or meds, ¿sometimes it does not allow you to,¿ and they ¿have to disconnect and reconnect several times before we can do iv push.¿ medline received 1 case of iv extension set; 10 were selected randomly for testing, visual inspection found no noticeable defects nor any cosmetic damage, and flushing was attempted with 10/10 succeeded in this function.The complaint was not confirmed.No further action was taken because objectionable trends were not seen for this issue, and the division has been informed and will continue to monitor trends.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|