| Model Number |
7840 |
| Medical Device Problem Codes |
Failure to Sense (1559); Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/12/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that no sensing activity was observed on this right atrial (ra) lead.Technical services (ts) was consulted and upon review of data determined that the lead was likely dislodged.Ts recommended an x ray to confirm lead location.No adverse patient effects were reported.This ra lead remains in service.
|
| |
|
Additional Manufacturer Narrative
|
|
This product investigation is still in progress.This report will be updated when product investigation is complete.If pertinent information is provided in the future, a supplemental report will be submitted.
|
| |
|