The customer reported error 1 message and stated that the item is not at room temp.Medline received 1 instant read digital temple thermometer in used condition with 2 aaa working batteries, tested the sample according to the instructions for use, sanitized the thermometer¿s probe sensor and attempted to test the sample several more times, and replaced and inserted 2 new aaa batteries.The complaint was confirmed; no temperature readings were displayed, the thermometer would only display an "er 1" message, and replacing the batteries did not solve the issue.Issue shown via investigation was caused due to a functional malfunction, and the sample did not work as intended.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|