Model Number 8637-20
Medical Device Problem Codes Display or Visual Feedback Problem (1184); Charging Problem (2892)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/06/2026
Type of Reportable Event Malfunction
Event or Problem Description
Information was received from a consumer regarding a patient with an implantable pump containing unknown drug for non-malignant pain.It was reported that the patient's communicator was not charging, they patient (pt) contacted their doctor but was told to call patient services.Pt said the issue started about 6 months ago.Pt already tried to use different charger and outlet, but the communicator still won't charge.Pt confirmed no visible damage on the port of the communicator.When they powered on the communicator, they got the blue flashing light but no orange light on the battery indicator when plugged in to the charging cord.Pt stated that the same charging cord that they were using for the handset and communicator was working to their other device.Troubleshooting steps taken on the call didn't resolve the issue.Agent sent a request to replace the communicator.
 
Additional Manufacturer Narrative
Section d information references the main component of the system.Other relevant device(s) are: product id: tm90t0, serial/lot #: (b)(6), ubd: 10-jan-2025, udi#: (b)(4) h3: analysis of the communicator, model tm90t0, s/n (b)(6), revealed the micro usb charge port was loose.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCHROMED II
Common Device Name
PUMP, INFUSION, IMPLANTED, PROGRAMMABLE
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
justin ellis
8200 coral sea street ne mvs21
mounds view, MN 55112
+1(763)526-5677
MDR Report Key50898556
Report Number3004209178-2026-16327
Device Sequence Number5853051
Product Code LKK
UDI-Device Identifier00763000634094
UDI-Public(01)00763000634094(17)250828
Combination Product (Y/N)N
Initial Reporter StateOK
Initial Reporter CountryUS
PMA/510(K) Number
P860004
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 09/25/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/28/2025
Device Model Number8637-20
Device Catalogue Number8637-20
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/21/2026
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/23/2026
Initial Report FDA Received Date09/25/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/13/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
"See H11...."
Patient Age71 YR
Patient SexFemale
-
-