| Model Number |
TCDS0801-XTW |
| Medical Device Problem Codes |
Positioning Failure (1158); Entrapment of Device (1212)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
08/31/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2026 a patient presented with grade 5 functional tricuspid regurgitation (tr) for a triclip procedure.During the procedure, while trying to grasp the leaflets first triclip, it was unable to grasp both the leaflets.It was observed that lateral leaflet was caught between the gripper and the shaft of clip delivery system (cds).Multiple maneuvers were attempted to freed the clip but was unsuccessful.Clip was deployed on chordae and lateral leaflet.Additional clip was implanted to stable the first clip.All steps were performed as per ifu.Imaging was difficult.The final tr was grade 3.There were no adverse patient effects or clinically significant delays reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
| |
|