Information was received from a patient regarding an external device to an implantable pump infusing hydromorphone, bupivacaine, and fentanyl for non-malignant pain.It was reported that their communicator was not holding a charge very long.Patient stated they just charged it yesterday and it was already down to 44%.Patient stated this had been going on for about 2 weeks now.A request was sent to the repair department.
|
|
D9.The concomitant product tm90 (communicator) was returned on sep-16-2026 and available for evaluation.D10: section d information references the main component of the system.Other relevant device(s) are: product id: tm90t0, serial #: (b)(6), ubd: 15-aug-2024, udi#: (b)(4), product type: accessory h3.Analysis of the communicator identified the micro usb charge port was loose.The device passed battery charging bench test, passed final functional jbal test, and was taken out of service/scrapped.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|