The patient reported to the nurse that he connects the bag, then when he gets up the urine drips out of the emptying valve without the valve being open.Medline quality received 3 ea of the reported product; the product was received in packaged, unused condition, visual assessment revealed no defects nor abnormalities, functional testing showed the valve functioned as normal with no signs of leakage, and a hang test revealed no defects nor abnormalities.The reported issue of twist-valve leakage could not be verified based on visual or functional inspection of the received sample, and the cause was not confirmed with the received sample.There are currently no known trends for this product lot.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|