U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Medline Industries, LP MEDLINE; BAG,LEG,TWIST-VALVE,STRAPS,MEDIUM,20OZ

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Medline Industries, LP MEDLINE; BAG,LEG,TWIST-VALVE,STRAPS,MEDIUM,20OZ Back to Search Results
Catalog Number DYND12574H
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/30/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that "he connects the bag, then when he gets up the urine drips out of the emptying valve without the valve being open.".
 
Additional Manufacturer Narrative
The patient reported to the nurse that he connects the bag, then when he gets up the urine drips out of the emptying valve without the valve being open.Medline quality received 3 ea of the reported product; the product was received in packaged, unused condition, visual assessment revealed no defects nor abnormalities, functional testing showed the valve functioned as normal with no signs of leakage, and a hang test revealed no defects nor abnormalities.The reported issue of twist-valve leakage could not be verified based on visual or functional inspection of the received sample, and the cause was not confirmed with the received sample.There are currently no known trends for this product lot.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
BAG,LEG,TWIST-VALVE,STRAPS,MEDIUM,20OZ
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
vamshi thupakula
three lakes drive
northfield, IL 60093--275
MDR Report Key50898620
Report Number1417592-2026-51022
Device Sequence Number10761485
Product Code KNX
UDI-Device Identifier10080196990337
UDI-Public10080196990337
Combination Product (Y/N)N
Initial Reporter StateOK
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/24/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberDYND12574H
Device Lot Number96924070001
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer02/10/2025
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/31/2025
Initial Report FDA Received Date09/25/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-