• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Alcon Laboratories Ireland Ltd. Clareon Toric Aspheric Hydrophobic Acrylic IOL with AutonoME Delivery System; Lens, Intraocular, Toric Optics

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Alcon Laboratories Ireland Ltd. Clareon Toric Aspheric Hydrophobic Acrylic IOL with AutonoME Delivery System; Lens, Intraocular, Toric Optics Back to Search Results
Model Number CNA0T5
Medical Device Problem Code Device Damaged by Another Device (2915)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 08/25/2026
Type of Reportable Event Malfunction
Event or Problem Description
A healthcare professional reported that during an intraocular lens implant procedure when inserting into the eye, the haptic was broken between the plunger, so the intraocular lens was replaced.The lens was replaced with unspecified lens during the same procedure.Additional information has been received from surgeon, stated that the trailing haptic appeared to be slightly tucked prior to insertion.Because of concerns about advancing the relatively long plunger too far into the eye, the surgeon stopped advancing the plunger midway.Upon resuming advancement, the trailing haptic had fully unfolded.Subsequently, the fully unfolded trailing haptic became trapped between the plunger and cartridge and was torn off during insertion.
 
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Clareon Toric Aspheric Hydrophobic Acrylic IOL with AutonoME Delivery System
Common Device Name
Lens, Intraocular, Toric Optics
Manufacturer (Section D)
Alcon Laboratories Ireland Ltd.
cork business&technology park
model farm road
cork 00000
EI  00000
Manufacturer (Section G)
Alcon Laboratories Ireland Ltd.
cork business&technology park
model farm road
cork 00000
EI   00000
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
+1(800)757-9780
MDR Report Key50898726
Report Number9612169-2026-02197
Device Sequence Number14910144
Product Code MJP
UDI-Device Identifier00380652399753
UDI-Public(01)00380652399753(17)280910(11)240912(10)25920092
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P190018
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/25/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCNA0T5
Device Lot Number25920092
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Received by Manufacturer 08/31/2026
Initial Report FDA Received Date09/25/2026
Date Device Manufactured09/12/2024
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PROVISC OVD
Patient SexUnknown
-
-