On 11 sep 2026, it was reported by an arthrex employee via email that an ar-1588rt-j acl tightrope rt w/ deploying suture's suture failed.Upon inserting the graft and confirming the button flip, the itb was found to be protruding.An additional incision was made and pushed back into place using an instrument such as a pean clamp.Subsequently, the tensioning suture failed during the adjustment phase; however, since approximately 10 cm of the suture remained, adjustment was still possible
the procedure was completed successfully using the same product.No patient harm was reported.
|