• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Medline Industries, LP MEDLINE; ANEROID,HAND HELD,ADULT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Medline Industries, LP MEDLINE; ANEROID,HAND HELD,ADULT Back to Search Results
Catalog Number MDS9380
Medical Device Problem Codes Gas/Air Leak (2946); Material Split, Cut or Torn (4008); Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/30/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that "the black hand pump to inflate the cuff is no longer reinflating when releasing pressure on it.".
 
Additional Manufacturer Narrative
The customer reported "the black hand pump to inflate the cuff is no longer reinflating when releasing pressure on it." we received 1ea under the complaint of "no longer inflating," and upon initial inspection noticed that the bulb would not release when pressed; the bottom of the bulb was cleared where the bearing was stuck, and another issue was discovered, a leakage in the bladder due to a small tear.The complaint is confirmed under leakage.Cause was documented as "comp./part defective, broke," and there was no reported death or serious injury associated with this event.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
ANEROID,HAND HELD,ADULT
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
konstantina callas
three lakes drive
northfield, IL 60093--275
MDR Report Key50898733
Report Number1417592-2026-51030
Device Sequence Number5829464
Product Code DXQ
UDI-Device Identifier10080196717743
UDI-Public10080196717743
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K211084
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/24/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberMDS9380
Device Lot Number65122040002
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer01/06/2026
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/30/2025
Initial Report FDA Received Date09/25/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-