The customer reported "the black hand pump to inflate the cuff is no longer reinflating when releasing pressure on it." we received 1ea under the complaint of "no longer inflating," and upon initial inspection noticed that the bulb would not release when pressed; the bottom of the bulb was cleared where the bearing was stuck, and another issue was discovered, a leakage in the bladder due to a small tear.The complaint is confirmed under leakage.Cause was documented as "comp./part defective, broke," and there was no reported death or serious injury associated with this event.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|