According to the customer, during a paracentesis procedure, the flow of fluid from the patient to the vacutainer was impacted when the tubing kinked, and efforts failed to restore the flow of fluid; the customer reported there was no serious injury, medical intervention, or follow-up care related to the reported incident.A photo and a sample were received for investigation, including three photos showing various kinks along the tubing and one used roller extension set with the tubing; there were multiple kinks within the tubing, the roller did not create any additional kinks, and the kinks present in the tubing did not straighten when the tubing was pulled.The vendor reviewed the sample provided and confirmed the reported issue.The vendor determined the kink was not caused by the use of the roller clamp, however they were unable to determine a true root cause; the complaint was entered into the vendor database for monitoring and trending, and no further action was taken by the vendor.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
|