• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Medline Industries, LP MEDLINE; TOWEL,OR,DSP,ST,BLUE,STD,4/PK,20PK/CS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Medline Industries, LP MEDLINE; TOWEL,OR,DSP,ST,BLUE,STD,4/PK,20PK/CS Back to Search Results
Catalog Number MDT2168284
Medical Device Problem Code Particulates (1451)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/02/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that "the towels are linting with wire use in the ir.".
 
Additional Manufacturer Narrative
A complaint was received related to linting from or towels, and the reporter stated the towels were linting with wire use in the ir.The received samples were evaluated, the vendor was notified of the issue, and trend analysis was performed by reviewing the past two years of complaint data; approximately 10 or towels were received in a single bag without the original packaging and in a wet/damp state.In the condition that the towels were received, it was difficult to confirm the reported issue, and the reported issue was not confirmed.The vendor noted that there was nothing found within the relevant records and retained samples that indicate potential causes to the reported issue, and the investigation indicated no need for further action at this time.There were no reports of death, serious injury, medical intervention, or follow up care.It has been determined that the reported event could cause or contribute to serious injury if it were to occur again.In an abundance of caution, this is a reportable event.If additional information becomes available this report will be reopened and reevaluated.This medical device report (mdr) is being submitted as part of retrospective review activities, which include review of historical data in accordance with regulations at 21 cfr part 803.The information included in this mdr reflects the information reasonably known to the manufacturer at the time of this retrospective review and submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDLINE
Common Device Name
TOWEL,OR,DSP,ST,BLUE,STD,4/PK,20PK/CS
Manufacturer (Section D)
Medline Industries, LP
three lakes drive
northfield IL 60093 -275
Manufacturer Contact
rahul kota
three lakes drive
northfield, IL 60093--275
MDR Report Key50898795
Report Number1417592-2026-51045
Device Sequence Number6001087
Product Code FRL
UDI-Device Identifier40080196282730
UDI-Public40080196282730
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Type of Report Initial
Report Date (Section B) 09/25/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberMDT2168284
Device Lot Number6052405019
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer01/09/2025
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/02/2025
Initial Report FDA Received Date09/25/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1